Abstract
The pulmonary artery (PA) catheter has been used to determine hemodynamic indices; however, it has recently been criticized. This study was undertaken to evaluate an esophageal Doppler monitor (EDM) as a possible replacement for PA catheter in critically ill, mechanically ventilated patients. EDM and PA catheters were placed in patients in the surgical intensive care units (n = 14, 118 matched sets of data). PA catheter and EDM measurements, including corrected flow time (FTc,) a measure of preload, were obtained. Pearson correlation (r) was analyzed to compare PA catheter and EDM measurements, and a nonlinear regression model was used to describe Starling Relationships. Cardiac output correlated between EDM and PA catheter (r = 0.6; p < 0.001). FTc correlated more strongly with cardiac output than did pulmonary capillary wedge pressure. (FTc: r2 = 0.27; p < 0.001; cardiac output: r2 = 0.04; p = 0.06). Corrected flow time is a better indicator of preload than pulmonary capillary wedge pressures. EDM seems to be at least as useful as PA catheter in managing the hemodynamic status of critically ill surgical patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The Journal of Trauma: Injury, Infection, and Critical Care
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.