Abstract

Skin erythemas formed in three patients during surgery at the sites where negative electrodes had been attached to stimulate the ulnar nerve for a neuromuscular transmission monitor (Relaxograph). The patients were all women, aged 52, 62, and 74 years, and general anesthesia lasted 8 h 20 min, 4 h 50 min, and 8 h 45 min, respectively. The electrodes used were disposable ECG electrodes in the first two patients and one designed for a neuromuscular monitor in the third; all were carbon-coated and then covered with gel. However, when the electrodes were detached from the lesion, they all showed loss or damage of the carbon coating under the gel. We recommend balancing the merit of monitoring with the risk of complications, even when applying an apparently safe, noninvasive monitor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.