Abstract

We have studied 219353 individual clinical chemistry results obtained in methods comparison studies. Each result was prospectively compared with its replicate, comparative, or repeat value to identify differences from expected values. Unacceptable results were defined as differing from the expected values by < or = 7 SDs or CVs. We believe these differences represent special-cause variation and should be expressed as unacceptable rates per million results (ppm). We observed 447 ppm unacceptables: 196 ppm in control samples and 251 ppm in patients' samples. Results judged likely to alter patient care occurred at a rate of 41 ppm. To better understand the magnitude of these rates, we compared these results with reports of error rates in HIV testing and the airline industry. The measurements reported were made for the purpose of quality improvement, not judgment or discovery. The significance of these findings for laboratorians, manufacturers, and regulators is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.