Abstract

This article experimentally investigates and assesses the errors that may be incurred in the hue-based thermochromic liquid crystal thermochromic liquid crystal (TLC) method, and their causes. The errors include response time, hysteresis, aging, surrounding illumination disturbance, direct illumination and viewing angle, amount of light into the camera, TLC thickness, digital resolution of the image conversion system, and measurement noise. Some of the main conclusions are that: (1) The 3×8 bits digital representation of the red green and blue TLC color values produces a temperature measurement error of typically 1% of the TLC effective temperature range, (2) an eight-fold variation of the light intensity into the camera produced variations, which were not discernable from the digital resolution error, (3) this temperature depends on the TLC film thickness, and (4) thicker films are less susceptible to aging and thickness nonuniformities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.