Errors which may occur in the measurement of transverse relaxation rates in liquids by wideband multiple pulse techniques are described. Numerical analysis of Bloch equations is used to evaluate the effects of imperfect pulses, and results of pulse imperfections are demonstrated experimentally. Effects of field-frequency instability and of diffusion combined with external field gradients are discussed. Methods for avoiding several of these errors are described and applied to the measurement of T 2 for degassed benzene. It is experimentally demonstrated that if certain criteria are met, measurements of T 2 with accuracy of ±1–2% are feasible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.