Abstract

Yong-Nam PakDepartment of Chemistry Education, Korea National University of Education, Chung Buk 363-791, KoreaE-mail: pakyn@knue.ac.krReceived July 2, 2014, Accepted August 8, 2014Matrix-induced mass bias and its effect on the accuracy of isotope ratio measurements have been examined fora quadrupole-based inductively coupled plasma-mass spectrometer (Q ICP-MS). Matrix-induced mass biaseffect was dire ctly proportional to % mass d ifference, and its magnitude varied for element and nebulizer flowrate. For a given element and conditions in a day, the effect was consistent. The isotope ratio of Cd106/Cd114under 200 µg g

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.