Abstract

Three-dimensional strain fields in articular cartilage subjected to compressive loading can be determined using a recently developed MRI-based cartilage deformation by tag registration technique. The objective of this study was to determine the experimental variables that minimize the technique error, which has not been previously reported. Error (strain bias and precision) was determined using direct experiments and Monte Carlo simulations for four variables: spatial resolution, tag line spacing, applied nominal strain, and number of control points used to describe tag lines in a B-spline model. The important results include the following: (1) bias was not significantly different from zero, (2) precision increased with image resolution and with tag line spacing, (3) precision was independent of applied nominal strain, and (4) error was a minimum (absolute precision = 0.41% strain) for the following values: spatial resolution = 0.05 x 0.05 mm2; tag line spacing = 2.0 mm; control points = 6. With these results the technique can now be used in various applications while minimizing error.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.