Abstract
The correction accuracy of an external fixator is crucial to the treatment outcome of deformity correction and patient safety. In this study, the mapping model is established between the pose error and kinematic parameter error of the motor-driven parallel external fixator (MD-PEF). Subsequently, the kinematic parameter identification and error compensation algorithm of the external fixator is established based on the least squares method. An experimental platform based on the developed MD-PEF and Vicon motion capture system is constructed for kinematic calibration experiments. Experimental results show that the correction accuracy of the MD-PEF after calibration is as follows: translation accuracy dE1 = 0.36 mm, axial translation accuracy dE2 = 0.25 mm, angulation accuracy dE3 = 0.27°, and rotation accuracy dE4 = 0.2°. The accuracy detection experiment verifies the kinematic calibration results, which further validates the feasibility and reliability of the error identification and compensation algorithm constructed by the least squares method. The calibration approach used in this work also provides an effective way to improve the accuracy of other medical robots.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.