Abstract

The errors are considered for an automated time-domain network analyzer based on a Tektronix sampling oscilloscope. Procedures are presented for removing the errors produced by two important sources: unwanted reflections associated with discontinuities and nonlinearities in the time base of the oscilloscope. The effectiveness of the error correction procedures is demonstrated by measurements of standard terminations with known reflection coefficients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.