Abstract

The current method for evaluating the effective dielectric constant εe f of microstripline (or coaxial line) has been found to be error susceptive. The error analysis has shown the limitations on the bandwidth of measurements and physical restrictions to be imposed on the test jigs to avoid anomalous values of propagation constant. The practical measurement of εe f has been carried further to characterise OSM–244–4ASF connectors and HI–Q 100pF DC block chip capacitor in the 2–4 GHz range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.