Abstract

The positional, electronic, and density accuracy of the Syntex AD-1 flatbed autodensitometer is discussed. The standard deviation of optical density measurements due to electronic noise is ± 0.001 optical density units at an optical density of 0, rising to ± 0.04 at an optical density of 2. No deviation from linearity in optical density measurements could be measured. During scanning, the standard deviation in position is less than 1.5 μm. Film grain noise for various films is compared and the effect of signal averaging on this noise is studied. This scanner is compared with other scanners available for accuracy of data collection from precession films.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.