Abstract

We recently treated a patient in whom a Gore TAG thoracic endograft (W.L. Gore and Assoc, Flagstaff, Arix) had been used to repair a descending thoracic aneurysm as the second stage of a hybrid procedure. This patient had previously undergone repair of ascending and aortic arch aneurysms, with an elephant trunk graft limb placed in the descending thoracic aorta for subsequent repair of the descending thoracic aneurysm. Eight months after placement of the thoracic endograft, the patient presented with an acutely expanding and symptomatic thoracic aneurysm. The patient was operated on urgently. The proximal portion of the endograft had eroded into the previously placed Dacron elephant trunk limb. The proximal portion of the endograft was removed and was replaced with a Dacron graft. The management of this patient forms the basis of this report.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.