Abstract

The results of a follow-up study concerning the decontamination of Mycobacterium chelonae subspecies abscessus from the bronchofibrescopes and the automated bronchoscope disinfection machine are described in this paper. After modification of the methods for disinfecting the bronchofibrescopes (adding a disinfection procedure with 70% alcohol before using the automated bronchoscope disinfection machine, increasing glutaraldehyde concentration to 3%, and changing the glutaraldehyde solution once a week), and the automated bronchoscope disinfection machine (recirculating used disinfectant), M. chelonae has not been detected from either the bronchofibrescopes or the automated bronchoscope disinfection machine (examined every 6 months for 4 yr by microscopy and cultures). Moreover, no M. chelonae has been clinically detected from bronchial washings for 4 yr.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.