Abstract

Using standard laboratory equipment a statistical comparison was carried out of several photometric methods (normal and various Higuchi plots) and potentiometric methods (point of inflection and Gran) for equivalence-point location in an acetic acid-sodium hydroxide solution titration. All methods exhibited similar precision but were divided into two groups with respect to accuracy. With increasing carbonate concentration in the titrant there was a decrease in the linear range for Higuchi and Gran plots and a decrease in accuracy for all methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.