Abstract
The number of fluoroscopy and fluoroscopically guided procedures has been substantially growing in developing countries at the same time advanced and sophisticated equipment are used in some hospitals. However, radiation protection requirements are not necessarily well adopted. In this study nine fluoroscopy X-ray units in Sudan were examined for compliance with international standards. The tests included: beam quality, entrance surface air kerma, image quality and radiation field measurements. Staff radiation protection tools such as lead aprons and eye glasses were also visually examined to find out whether international recommendations were fulfilled and to determine the level of staff awareness. The measured peak tube voltage deviation exceeded the recommended tolerance level in 30 % of the measurements. The results of patient doses measurements exceeded the recommended reference dose levels in 43 % of the measurements; however image quality and radiation field generally fulfilled the requirements for most units. The study revealed that a considerable number of fluoroscopy units were not performing according to the international standards and highlights the need of optimisation of radiation protection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.