Abstract

In this study, a paper-based sensor combined with visual distance-readout technique for point of-care testing (POCT) of urea was developed by urease-mediated chitosan viscosity change. A series of factors that affect the performance of the sensor were investigated, including the type of filter paper, chitosan concentration, acetic acid concentration and enzymatic reaction conditions. Under optimal conditions, the proposed method for urea determination has good linearity between 3.8–15.1 mM. The limit of quantitation is 3.8 mM. Finally, the paper-based sensor was successfully applied to the determination of urea in two diesel exhaust fluid (DEF) samples. The recoveries of urea were 91.4 % and 109.9 % in DEF-1 and DEF-2, respectively. The present study provides a novel approach, which integrates paper-based sensor and visual distance-readout technique, for monitoring urea in POCT application, especially in remote or resource-limited regions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.