Abstract

We developed an equilibrator inlet-proton transfer reaction-mass spectrometry (EI-PTR-MS) method for fast detection of dimethyl sulfide (DMS) dissolved in seawater. Dissolved DMS extracted by bubbling pure nitrogen through the sample was continuously directed to the PTR-MS instrument. The equilibration of DMS between seawater and the carrier gas, and the response time of the system, were evaluated in the laboratory. DMS reached equilibrium with an overall response time of 1 min. The detection limit (50 pmol L(-1) at 5 s integration) was sufficient for detection of DMS concentrations in the open ocean. The EI-PTR-MS instrument was deployed during a research cruise in the western North Pacific Ocean. Comparison of the EI-PTR-MS results with results obtained by means of membrane tube equilibrator-gas chromatography/mass spectrometry agreed reasonably well on average (R(2) = 0.99). EI-PTR-MS captured temporal variations of dissolved DMS concentrations, including elevated peaks associated with patches of high biogenic activity. These results demonstrate that the EI-PTR-MS technique was effective for highly time-resolved measurements of DMS in the open ocean. Further measurements will improve our understanding of the biogeochemical mechanisms of the production, consumption, and distribution of DMS on the ocean surface and, hence, the air-sea flux of DMS, which is a climatically important species.