Abstract

Task Group 21 of the American Association of Physicists in Medicine defined the cavity-gas calibration factor Ngas for ionization chambers in a protocol for radiotherapy dosimetry published in this journal [Med. Phys. 10, 741 (1983)]. Later, Schulz et al., [Med. Phys. 13, 755 (1986)] published a letter of clarification that proposed a revised equation for Ngas in which the effect of atmospheric humidity was specifically dealt with. The present report points out errors in both of those presentations of the Ngas equation, and derives formulations for cases where the standardization laboratory does or does not correct for ambient humidity during chamber calibration. In the case where humidity is corrected for, the resulting quantity is more properly called "Nair." The Ngas equation is stated in terms of the air kerma calibration factor Nk as well as the more familiar exposure calibration factor Nx.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.