Abstract

The OSL dosimetry technique with teeth has been validated against the EPR dosimetry technique through a measurement comparison performed on the same teeth. The OSL reconstructed doses were found to be in agreement with corresponding EPR doses. Minimum measurable doses for the OSL technique were estimated to be in the range 0.9–1.5 Gy for measurements made within 24 h post-exposure if OSL signals are collected from 6 teeth at the same time. These values satisfy the requirements for emergency triage dosimetry. The fading of tooth OSL signals correlated with fading of radiation-induced EPR signals observed at g = 2.0115 that are attributed to CO3- radicals. OSL sensitivity can be enhanced if more teeth from the same individual will be used for the signal accumulation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.