Abstract
The ePix100 x-ray camera is a new system designed and built at SLAC for experiments at the Linac Coherent Light Source (LCLS). The camera is the first member of a family of detectors built around a single hardware and software platform, supporting a variety of front-end chips. With a readout speed of 120 Hz, matching the LCLS repetition rate, a noise lower than 80 e-rms and pixels of 50 µm × 50 µm, this camera offers a viable alternative to fast readout, direct conversion, scientific CCDs in imaging mode. The detector, designed for applications such as X-ray Photon Correlation Spectroscopy (XPCS) and wavelength dispersive X-ray Emission Spectroscopy (XES) in the energy range from 2 to 10 keV and above, comprises up to 0.5 Mpixels in a very compact form factor. In this paper, we report the performance of the camera during its first use at LCLS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.