Abstract

Abstract The Sector 7 undulator beamline (7 ID) of the Advanced Photon Source (APS) is dedicated to time-resolved X-ray research [1] . Silicon avalanche photodiodes (APDs) are used as the primary point detector for time-resolved Bragg diffraction experiments for their fast recovery time ( [2] . Common practice has been to attenuate the monochromatic beam entering the experimental hutch to an appropriately low flux [3] . For these high-flux experiments, an APD operated in proportional mode is a better detector choice due to a large dynamic range and linearity. With the ZT4212 ZTEC, EPICS based oscilloscope, the operating procedure to use an APD in proportional mode has been improved. This article shows the setup and operating procedure for this oscilloscope and demonstrates its application to measuring time-resolved rocking curves of laser excited semiconductors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.