Abstract
A total of 15 patients plans and 79 DCA generated in iPlan RT Dose 4.1.2 (BrainLAB, AG, Germany), with a Pencil Beam algorithm using a kernel resolution of 2.5mm were tested. Mean equivalent MLC field size was of 16.6 mm (range 7.5-32.7mm). Plans were delivered on a Novalis TX, using the 6XSRS energy mode with a D/R of 1000 MU/min and a HDMLC multileaf collimator (2.5 mm leaf size at isocenter). Using a Lucy® 3D QA Phantom (Standard Imaging, Inc., USA), dose distributions were recalculated for each plan and Figure 2. In blue the Lucy phantom with the DD and in grey, bottom, with the PSD, position “tip down”. The water equivalence and absence of metal in the PSD resulted in no change in data for this detector. Both DD and PSD results were within the overall measurements accuracy. Analysis of the measurements performed with the detectors tip up and down, Fig.3, are shown below in Table 3.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.