Abstract

A sensitive spectrophotometric method for the determination of aspartame in beverages is described. The method involves the enzymic conversion of aspartame into formaldehyde by the alpha-chymotrypsin-alcohol oxidase system, followed by the formation of a chromophore with 4-aminopent-3-en-2-one. The calibration graph was linear in the range 2.0-30.0 micrograms ml-1 of aspartame. Many common ingredients of beverages do not interfere with the proposed method. The method was applied to the determination of the aspartame content of various real samples, and the results obtained were compared with those given by high-performance liquid chromatography.