Abstract
Manual procedures for estimating serum total cholesterol by use of cholesterol oxidase (EC 1.1.3.6) and the phenol--aminophenazone--peroxidase chromogenic system are described, in which cholesteryl esters are hydrolyzed either by use of pancreatic cholesterol ester hydrolase (EC 3.1.1.13) or saponification by ethanolic potassium hydroxide. Both methods are linear up to a cholesterol concentration of 12 mmol/L and are reproducible (between-run CV, about 1.1%). The chemical hydrolysis method yields results that are about 10% lower than those obtained by enzymic hydrolysis, because of incomplete removal of interfering thiols generated during the saponification of serum. The chemical hydrolysis procedure is much less susceptible to interferences, particularly by bilirubin, but the enzymic hydrolysis system is simpler to perform and therefore has a greater potential for mechanization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.