Abstract

Enzyme-amplified biochemical detection (EA-BCD) was used as a post-column detection technique, coupled on-line with high-performance liquid chromatography (HPLC). The enzyme detection system was developed to detect biotin or biotin containing compounds. Biotinylation is widely used to label analytes of interest ranging from small molecules to proteins and DNA. Naphthalene aldehyde and anthracene aldehyde were used as model compounds. Both compounds were biotinylated off-line with biotin aminocaproic hydrazide (BACH). On-column biotinylation was performed by preconcentration of anthracene aldehyde on copper phthalocyanine. After biotinylation, samples were introduced to the HPLC system. Enzyme-labeled streptavidin, which possesses high affinity to biotin, was added post-column to the HPLC effluent. Excess of enzyme-labeled affinity protein was removed by means of an immobilized biotin column. After separation of free and bound fraction, substrate was added, which was converted to a fluorescent product by the enzyme label. Using alkaline phosphatase as an enzyme label, a mass detection limit after on-column preconcentration and biotinylation of 250 fmol was achieved.