Abstract

Acrolein is a pollutant released daily to the indoor environment from different sources. The present study reports the development of a simple and sensitive cold fiber solid phase microextraction sampling method for the determination of acrolein in exhaled air and indoor air by gas chromatography mass spectrometry. O-(2,3,4,5,6-pentafluorobenzyl) hydroxylamine was used as derivatizing agent supported on a 65-μm polydimethylsiloxane-divinylbenzene SPME fiber. An acrolein permeation tube at 326.25ngmin-1 rate was used to generate gaseous standards. The method shows good results for main validation parameters. The limits of detection and quantification were 2.88 and 5.08μgm-3, respectively, for indoor analysis; and 2.40 and 3.79μgm-3, respectively, for exhaled air analysis. The precision showed standard deviation ranges from 6.00 to 8.00% for intra-assay analyses and from 8.00 to 10.00% for inter-assay analyses. After optimizing the conditions, analyses of real samples were performed on indoor environments contaminated by cigarette smoke, or heated oil, including pastry shops, restaurants, churros stands, and closed parking cars located in the city of Belo Horizonte, Brazil. Acrolein breaths of exposed people were also determined. A good Pearson correlation coefficient (r = 0.901) was observed between the concentration of acrolein in indoor air and exhaled air, allowing to propose acrolein breath as environmental exposure biomarker. Graphical Abstract Cold fiber solid phase microextraction gas chromatography/mass spectrometry.