Abstract

Endoluminal repair of an infrarenal saccular aortic aneurysm was undertaken with a modular endovascular stent graft in a 74-year-old man. After deployment of the main body and right iliac limb of the graft, an attempt was made to gain guide wire access to the open short left limb of the graft for placement of the left iliac prosthesis. Attempts to gain access to the left limb graft from the left femoral artery area were unsuccessful. Since we were unable to pass the wire into the iliac system or snare it from below, can nulation from the left brachial artery was attempted. This resulted in the guide wire coiling up within the aneurysm. Eventual conversion to an open procedure confirmed that the short open left limb of the modular graft had deployed such that the opening was completely impacted against the inferior wall of the saccular aneurysm, preventing successful passage of a guide wire. The distance from the aortic neck to the base of the aneurysm is generally not considered an important parameter whe...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.