Abstract

Ablation procedures in the left atrium for treatment of atrial fibrillation are becoming increasingly common. The procedure often involves placing one or two circular mapping catheters in the left atrium. Entrapment of an ablation catheter in the mitral valve during ablations of left-sided accessory pathways by the retrograde approach has been described in two earlier published reports. More recently, several reports describe similar entrapment of a mapping catheter. In a recently published review, however, only one case of unspecified valve damage was registered among 8745 atrial fibrillation procedures. The purpose of this study was to evaluate patients with entrapment. Retrospective analysis of electrophysiological results. We describe three patients with entrapment during ablations for atrial fibrillation. The entrapments occurred with three different operators at three different electrophysiological laboratories within 2 years. The complication described here may be more common than is widely appreciated. From our figures, we estimate the incidence of the complication to 0.9% (95% confidence interval, 0.2-2.5%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.