Abstract

Background: Cone-beam computed tomography (CBCT) has gained considerable use, while radiation burden remains an important concern. Aims: This study aimed at determining the entrance skin dose of the thyroid gland area through exposure with the normal, soft, and hard modes of Vatech (Pax-i) panoramic device as well as different field of views (FOVs) of Vatech (Pax-i 3D) CBCT system. Materials and Methods: Dose measurements were performed on a head-and-neck phantom by an ion chamber dosimeter. Panoramic imaging was done in three normal, soft, and hard modes, each entailing woman, man, and child options. CBCT examinations were performed with various FOVs, each having normal, soft, and hard modes as well as woman, man, and child choices. Doses obtained from different protocols of the two imaging modalities were compared by paired t-test. Results: All FOVs in CBCT resulted in greater radiation dose than panoramic in each of the normal, soft, and hard exposure modes (P