Abstract
A preconcentration method for subsequent determination of rare earth elements (REE) by X-ray fluorescence (XRF) spectrometry was developed. The method is based on using (o-[3,6-disulfo-2-hydroxy-1-naphthylazo]-benzenearsonic acid) (Thorin) as a complexing agent which is retained on a polyamide membrane by a chemofiltration process. The pH dependence of the chemofiltration of these metal ions on the membrane and other variables, such as flow-rate, contact time, kinetic of complex formation, etc. were determined. The membrane containing the chemofiltrate formed a thin film, which eliminated the interelemental effects when measured by XRF. The detection limits were 23, 23 and 49 ng/mL for Sm(III), Eu(III) and Gd(III), respectively. High enrichment factors were obtained. The method was successfully applied to the preconcentration of Sm(III), Eu(III) and Gd(III) from different samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.