Abstract
In chromatic confocal measurement (CCM), the accuracy and precision of the results rely on the robustness of the spectral power distribution (SPD) signal resolution. However, traditional centroid algorithms struggle with the inherent discretization of the SPD signal, leading to significant measurement errors. To address this issue, a linear interpolation based centroid algorithm is proposed to improve the robustness of the results. In the precision calibration experiment conducted with the in-house developed chromatic confocal sensors, the proposed algorithm demonstrates good robustness. Moreover, it achieves a 50% enhancement in the accuracy of a 4 mm range, improving from 1 μm to 0.5 μm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.