Abstract

Matrix-assisted laser desorption ionization time-of-flight (MALDI-TOF) mass spectrometers are simple and robust mass spectrometers used for analysis of biologically relevant molecules in diverse fields including pathogen identification, imaging mass spectrometry, and natural products chemistry. Despite high nominal resolution and accuracy, we have observed significant variability where 30–50% of individual replicate measurements have errors in excess of 5 parts-per-million, even when using 5-point internal calibration. Increasing the number of laser shots for each spectrum did not resolve this observed variability. What is responsible for our observed variation? Using a modern MALDI-TOF/TOF instrument, we evaluated contributions to variability. Our data suggest a major component of variability is binning of the raw flight time data by the electronics and clock speed of the analog-to-digital (AD) detection system, which requires interpolation by automated peak fitting algorithms and impacts both calibration and the observed mass spectrum. Importantly, the variation observed is predominantly normal in distribution, which implies multiple components contribute to the observed variation and suggests a method to mitigate this variability through spectrum averaging. Restarting the acquisition impacts each spectrum within the electronic error of the AD detector system and defines a new calibration function. Therefore, averaging multiple independent spectra and not a larger number of laser shots leverages this inherent binning error to mitigate variability in accurate MALDI-TOF mass measurements.

Highlights

  • Matrix assisted laser desorption time-of-flight (MALDI-TOF) mass spectrometers are highly robust and capable instruments for biomolecular analysis

  • While the first practical TOF mass spectrometer was developed in the 1950s[1], it was the advent of new ionization techniques compatible with large biological molecules like MALDI[2] and electrospray ionization[3] that revolutionized the biological applications of mass spectrometers including TOF instruments

  • Not all peptides were observed in each spectrum

Read more

Summary

Introduction

Matrix assisted laser desorption time-of-flight (MALDI-TOF) mass spectrometers are highly robust and capable instruments for biomolecular analysis. Spectrum Averaging Enhances MALDI Mass Spectrometry ionization techniques and their impact on vastly improved mass spectrometer instrumentation performance. Modern MALDI-TOF mass spectrometers use delayed extraction and ion reflector systems to enhance instrument resolution and accuracy, enabling accurate mass measurements of peptides and molecules (S1 Fig.)[15,16]. Routine performance specifications for reflector MALDI-TOF instruments often exceed 15,000 for resolution measured by full-width at half-maximum (FWHM) and

Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.