Abstract
A series of system enhancements for the Josephson array voltage standard (JAVS) have been implemented to improve the efficiency and quality of dc voltage dissemination in the National Metrology Centre, A*STAR. Periodic direct comparison between two JAVS systems has been adopted as a quality assurance procedure to ensure accurate reproduction of the volt for calibration of secondary standards. A miniature millimeter-wave synthesizer is employed for better control of microwave power and locking frequency. Low thermal terminal reversing switches are introduced to minimize zero-offset measurement uncertainty. Monte Carlo method and ANOVA method are implemented in the data analysis and reporting.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.