Abstract

The effects of varying the concentrations of cross-linker N, N ′ -methyelene-bis-acrylamide (BIS) from 2% to 4%, and 2-hydroxyethylacrylate (HEA) monomer from 2% to 4% at 5% gelatin on the dose response of BIS–HEA–gelatin (BHEAG) aqueous polymer gel dosimeters have been studied using magnetic resonance imaging (MRI) for relaxation rate ( R 2 ) of water proton. The dosimeters were irradiated with 60Co teletherapy γ -ray source at a constant dose rate, receiving doses up to 30 Gy. The radiation polymerization occurs and increases with increasing initial dose. R 2 is found to decrease mono-exponentially with depth inside the polymer gel and depend strongly upon the initial concentrations of co-monomers (HEA and BIS). Dose–depth map for BHEAG gel was determined for different concentrations of co-monomer (HEA and BIS). The percentage dose depth was also evaluated which leads to a good agreement with the ionization chamber measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.