Abstract

To develop and demonstrate the feasibility of a new technique for respiratory-gated, fat-suppressed, three-dimensional spoiled gradient-recalled echo (3D-SPGR) with navigator gating for more accurate and robust motion detection. A navigator-gated 3D-SPGR technique was modified to include a wait period immediately prior to the navigator sequence for magnetization recovery. Furthermore, a variable flip angle scheme was realized by a combination of ramp-up, ramp-down, and attenuation strategies for optimizing point spread functions. Phantom and human experiments were conducted with our technique on 1.5T scanners. Using the method, T1-weighted 3D images with improved signal homogeneity were acquired with a maximum flip angle of 30° in phantom and human tests. Also, compared with the conventional navigator-gated 3D-SPGR, accurate respiratory motion detection of free-breathing subjects was provided, leading to reduced motion artifacts. The combination of wait insertion and the variable flip angle method improved both motion detection accuracy and image homogeneity in a navigator-gated 3D-SPGR study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.