Abstract

A histogram-based denoising algorithm was developed to effectively reduce ghost artifact noise and enhance the quality of an optical coherence tomography (OCT) imaging system used to guide surgical instruments. The noise signal is iteratively detected by comparing the histogram of the ensemble average of all A-scans, and the ghost artifacts included in the noisy signal are removed separately from the raw signals using the polynomial curve fitting method. The devised algorithm was simulated with various noisy OCT images, and >87% of the ghost artifact noise was removed despite different locations. Our results show the feasibility of selectively and effectively removing ghost artifact noise.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.