Abstract
A compact multi-bounce attenuated total reflection (ATR) probe combined with a Fabry-Pérot filter spectrometer (FPFS) has been developed for detection of hydrogen peroxide used for oxidative gas scrubbing operating in the mid-infrared (MIR) spectral region. A novel MIR supercontinuum light source is employed to enhance the quantification capabilities of the sensor and is compared to a classical thermal emitter. An improvement of a factor of 4 in noise and approximately a factor of 3 in limit of detection is shown in this study allowing fast inline detection of aqueous hydrogen peroxide solutions around 0.1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.