Abstract

Stent implantation is a principal therapeutic approach for coronary artery diseases. Nonetheless, the presence of stents significantly interferes with in-stent luminal (ISL) visualization and complicates the diagnosis of in-stent restenosis (ISR), thereby increasing the risk of misdiagnoses and underdiagnoses in coronary computed tomography angiography (CCTA). Dual-energy (DE) CT could calculate the volume fraction for voxels from low- and high-energy images (LHEI) and provide information on specific three basic materials. In this study, the innovative coronary stent decomposition algorithm (CSDA) was developed from the DECT three materials decomposition (TMD), through spectral simulation to determine the scan and attenuation coefficient for the stent, and preliminary execution for an in vitro sophisticated polyether ether ketone (PEEK) 3D-printed right coronary artery (RCA) replica. Furthermore, the whole-coronary-artery replica with multi-stent implantation, the RCA replica with mimetic plaque embedded, and two patients with stent further validated the effectiveness of CSDA. Post-CSDA images manifested no weakened attenuation values, no elevated noise values, and maintained anatomical integrity in the coronary lumen. The stents were effectively removed, allowing for the ISL and ISR to be clearly visualized with a discrepancy in diameters within 10%. We believe that CSDA presents a promising solution for enhancing CCTA diagnostic accuracy post-stent implantation.