Abstract

Many biologically relevant glycoproteins need to be separated on 1D- or 2D-gels prior to analysis and are available in picomole amounts. Therefore, it is important to have optimized methods to unravel the glycome that combine in-gel digestions with MALDI-TOF-MS. In this technical report, we investigated how the detection of in-gel released N-glycans could be improved by MALDI-TOF-MS. First, an AnchorChip target was tested and compared to ground steel target using several reference oligosaccharides. The highest signals were obtained with an AnchorChip target and D-arabinosazone as the matrix; a LOD of 1.3 to 10 fmol was attained. Then, the effect of octyl-β-glucopyranoside, a nonionic detergent, was studied during in-gel peptide-N(4) -(acetyl-ß-glucosaminyl) asparagine amidase F digestion of standard glycoproteins and during glycan extraction. Octyl-β-glucopyranoside increased the intensity and the amount of detected neutral as well as acidic N-glycans. A LOD of under 7 pmol glycoprotein could be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.