Abstract

An energy-discriminating K-edge x-ray Computed Tomography (CT) system is useful for increasing contrast resolution of a target region and for diagnosing cancers utilizing a drug delivery system. The CT system is of the first generation type and consists of an x-ray generator, a turn table, a translation stage, a two-stage controller, a cadmium telluride (CdTe) detector, a charge amplifier, a shaping amplifier, a multi-channel analyzer (MCA), a counter board (CB), and a personal computer (PC). The K-edge CT is accomplished by repeating translation and rotation of an object. Penetrating x-ray spectra from the object are measured by a spectrometer utilizing the CdTe detector, amplifiers, and MCA. Both the photon energy and the energy width are selected by the MCA for discriminating photon energy. Enhanced iodine K-edge x-ray CT was performed by selecting photons with energies just beyond iodine K-edge energy of 33.2 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.