Abstract

Introduction: In Yttrium-90 SPECT imaging, the energy window and collimator used during acquisition can have a major effect on image quality. In this work, we used a new and independent method to verify the prevoius results suggest different energy ranges, but all centered at 130 keV. Materials and method: We used Siemens Symbia SPECT-CT system fitted with High Energy General Purpose (HEGP), Medium Energy General Purpose (MEGP) and Low Energy High Resolution (LEHR) to acquire data from NEMA IEC PET Body Phantom with its 6 different spheres of 3.7, 2.8, 2.2, 1.7, 1.3, 1.0 cm diameter. Results:HEGP collimator is the most suitable for acquisitions of 90Y bremsstrahlung radiation from the point of view of the correct volume reproduction. For the bigger sphere’s study, the optimum ISO-counting curves is related to the energy range centred in 130 keV.Conclusion: The results obtained are consistent with previous studies. The Iso Counting Curves method can help to improve image quality.