Abstract

As different types of radionuclides (e.g., 131Cs source) are introduced for clinical use in brachytherapy, the question is raised regarding whether a relatively simple method exists for the derivation of values of the half value layer (HVL) or the tenth value layer (TVL). For the radionuclide that has been clinically used for years, such as 125I and 103Pd, the sources have been manufactured and marketed by several vendors with different designs and structures. Because of the nature of emission of low energy photons for these radionuclides, energy spectra of the sources are very dependent on their individual designs. Though values of the HVL or the TVL in certain commonly used shielding materials are relatively small for these low energy photon emitting sources, the question remains how the variations in energy spectra affect the HVL (or TVL) values and whether these values can be calculated with a relatively simple method. A more fundamental question is whether a method can be established to derive the HVL (TVL) values for any brachytherapy sources and for different materials in a relatively straightforward fashion. This study was undertaken to answer these questions. Based on energy spectra, a well established semiempirical mass attenuation coefficient computing scheme was utilized to derive the HVL (TVL) values of different materials for different types of brachytherapy sources. The method presented in this study may be useful to estimate HVL (TVL) values of different materials for brachytherapy sources of different designs and containing different radionuclides.