Abstract

Most textbooks on radiological equipment contain the information that the gain of an X-ray image intensifier is proportional to the ratio of the areas of its input and output phosphors, but the practical implications of this in terms of dose to the patient appear to be less well known. For the normal practice of performing fluoroscopy under automatic dose rate control, when switching to a magnified field of view the dose rate would be expected to increase in inverse proportion to the change in gain of the intensifier if a constant signal-to-noise ratio in the image is to be maintained. At the same time an automatic change in collimation should take place so that the radiation beam size is restricted to cover only the active portion of the intensifier input phosphor (NRPB, 1988).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.