Abstract

AbstractTrace element analyses by energy dispersive X-ray fluorescence measurements were made using synchrotron radiation from a dedicated electron storage ring at the Photon Factory in Japan. The continuum or the monochromatic beam was used for excitation. A crystal monochromator or two types of mirror systems were used for monochromatic excitation. A high signal to background ratio was attained with the crystal monochromator, while the highest absolute detectability was achieved with the mirror system. The minimum detection limita obtained from thin samples are of the order of 0.1 ppm or less than 0.1 pg. Furthermore the signal to background ratio was significantly improved by using an X-ray mirror as a sample support in which, external total reflection of exciting X-rays occured.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.