Abstract

This article presents a novel technique for measuring the concentration of calcite in copper ore samples using energy dispersive X-ray diffraction (ED-XRD). The results are compared with ultraviolet fluorescence spectroscopy (UVF) for the same group of samples. This approach substantially shortens the measuring time with respect to the current laboratory procedures from 300 min down to 15 min and 9 min for energy dispersive X-ray diffraction and ultraviolet fluorescence spectroscopy, respectively, while preserving a good accuracy level (RMSE < 0.577). The models used for data processing were calibrated and validated for both techniques via the correlation coefficients ( R c 2 and R v 2 ), the error of prediction (RMSEP) and the error of calibration (RMSEC). R v 2 values between 0.88 and 0.94 were obtained for UVF and ED-XRD data, respectively. The calibration and validation were performed for calcite concentration ranges of 3.4–10.7% in the case of UVF and 2.9–7.6% for ED-XRD. The results reported show the potential of this method for on-line estimation of calcite concentration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.