Abstract
Indicates that the sensitivity of LiF rods expressed as thermoluminescent response per rad in water, is dependent on both energy and geometrical dimensions of the LiF dosimeter. The sensitivity per rad in water of extruded LiF rods varies approximately by 25% for electron beam energies ranging from 1 to 20 MeV. A model was developed which predicted the change of sensitivity per rad in water of the LiF rods and the calculated and measured responses were compared.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.