Abstract

The responses of three types of silver phosphate glass dosimeter to X- and γ-radiation show a small energy dependence effect over the energy region, 1·26 to 24 MeV. Dosimeters in the form of thin glass plates exhibit a greater response to low energy X-rays. Increasing the size of the glass element reduces this energy dependence. Other investigators have found that by placing suitable lead filters about the glass, the response of the dosimeters was reasonably energy independent over the range 100 keV to 1 MeV. It has, however, been found that the lead filters cause phosphate glass dosimeters to show a high response to X-rays of several MeV in energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.