Abstract

Ten subjects with neurological lesion levels between T4 and T9 have been evaluated in the laboratory at ORLAU using their orthosis in its mechanical form alone (without FES assist). Initially the oxygen consumption at rest was measured. Then the patients were asked to walk (orthotic gait) at their preferred constant steady speed for a period of 5 minutes. Min. VO2 was measured during the exercise. From these readings the energy cost and consumption of the patients' gait was calculated. Our results show that the ParaWalker allows more efficient walking than do bilateral long leg braces and a paraplegic 3 point gait. It also shows that higher speeds with lower energy cost walking is possible with a different orthotic approach for high or mid thoracic lesion paraplegics. We believe that the energy demands to walk with the ParaWalker are within sustainable limits.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.