Abstract

We report on a simple method to perform an absolute calibration of the magnetic beam analyzer of the reaccelerator ReA3 at the National Superconducting Cyclotron Laboratory. The method is based on the time of flight between two beam stoppers 7.65 m apart. Based on two independent time-of-flight measurements at three different beam energies, the beam analyzer magnet is calibrated with an accuracy of 0.12 %, corresponding to a beam energy accuracy of 0.24 %.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.