Abstract

In recent years, spectral X-ray imaging using photon counting detectors (PCDs) becomes a hot topic in the field. For a PCD, each individual incident photon with different energies is analyzed and assigned to different energy channels according to pre-set thresholds respectively. Thus, the data from an energy channel carry spectral information. Because of the influence of pulse pileup, charge sharing and other physical effects, energy calibration of PCDs is a piece of challenging work. Various energy calibration methods of PCDs have been researched. Most of them demand extensive work with additional sources or equipments. In this work, we propose a novel approach for energy calibration by using only one monochromatic source. We use iterative optimization method to fully excavate and utilize the data. The method requires fewer experiments than other common-used calibration methods. Moreover, the charge sharing effect is implicitly taken into account in this method which is an important factor in the calibration of pixel detectors. We validated our method with radioactive sources. The resulting energy spectrum matched well as expected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.